S90 ZAHN Signature Series Dip Viscosity Cup
|
|
| All Stainless
Steel
S90/Zahn Signature Dip Viscosity cups are an improved version of the time
honored Zahn Signature viscosity cups.
-
Calibration of the S90/Zahn cups match the earlier Zahn Signature cups.
-
Conversion formulas and tables relating cup efflux time in seconds to viscosity
in centistokes are identical for both series of cups.
-
S90/Zahn cups qualify for certification under ANSI/NCSL Z540-1 or
MIL-STD-45662A, as applicable.
-
S90/Zahn cups are calibrated with standard
-
"G" series oils traceable to the National Institute of Standards &
Technology. "G" series calibration oils are produced in accordance with ISO
9002. NOTE: Efflux time from these cups does not meet ASTM specifications, but
tables are available for converting S90/Zahn cup efflux time to EZ™ Zahn cup
efflux time which complies with ASTM D 4212.
Certificate of Calibration available at additional charge for all cups.
|
|
|
S90/Zahn
Cup
Part
#
|
Cup
#
|
Seconds
Range
|
Centistoke
Range
|
|
61000.001
|
2
|
31 to 60
|
15 to 78
|
|
61000.002
|
3
|
19 to 60
|
39 to 238
|
|
|
|
| Calibrated
EZ™ ZAHN (ASTM)
Dip Viscosity Cup |
|
|
3% GUARANTEED TOLERANCE.
COMPLIES WITH & EXCEEDS ASTM D 4212
-
Oils used to standardize EZ Cups are produced in accordance with ISO 9002:
1984; EN 29002: 1984; BS5750 Part 2: 1984; ANSI/ASQC Q92: 1984.
-
The EZ Cup formula for each cup of the series matches the applicable ASTM
formula in D 4212 at the recommended calibration level.
-
EZ Cups are compatible to ASTM D 1084 Viscosity of Adhesives and ASTM D 816
-
EZ Cup calibration is traceable to the National Institute of Standards and
Technology
-
Calibration and Certification procedures qualify under ANSI/NCSL Z540-1 or MIL-
STD- 45662A as applicable
-
The EZ Cup is not matched by any other cup of its type, either with respect to
the advantages listed above, in highest quality of workmanship or in continuing
quality control procedures.
-
All stainless steel cup and handle.
-
THE FINEST, MOST RELIABLE, CALIBRATED & DOCUMENTED CUP ON THE MARKET.
|
 |
|
|
EZ
Zahn Cup
Part
#
|
Cup
#
|
Seconds
Range
|
Centistoke
Sensitivity
|
Midrange Oil Number¹
|
Midrange Oil Number¹
|
|
61100.001
|
2
|
40 to 60
|
10 to 36
|
1.3 |
G-10/19 |
|
61100.002
|
3
|
20 to 60
|
19 to 156
|
3.3 |
G-60/117 |
|
|
|
CONTINUING IMPROVEMENTS
The new EZ Viscosity Cup Series is a logical and necessary outgrowth of the
standardization studies on the Zahn Signature and similar cups for the
promotion of this most popular type of viscosity measuring instrument as a
national and international standard. Not only has this cup been improved with
respect to earlier produced cups and designed to comply to the requirements of
ASTM D 4212 but, in addition, each EZ Cup is furnished with a table which
permits conversion between efflux time in seconds to the nearest tenth of a
second to viscosity in centistokes. This table is particularly useful in
determining efflux time in seconds when viscosity in centistokes is known.
For those users who require documented certification of their measuring
equipment, the EZ™ cups may be ordered, at an additional charge, with a
CERTIFICATE of CALIBRATION . This document contains not only information on
actual cup calibration with standard oils traceable to the National Institute
of Standards and Technology but in addition, this certification also complies
to conditions and procedures under the requirements of ANSI/NCSL Z54O-1 or
MIL-STD-45662A as applicable.
The EZ™ cups are produced to very close mechanical tolerance in elaborate jigs
and fixtures. Such equipment not only insures that each cup is correctly
produced but also that all cups are identical. All parts of the cup are of
stainless steel except for the name plate. The following table provides
operating range specifications, cup midrange sensitivity and recommended
calibrating oils. Following this table are graphs and mathematical formulas
that relate efflux time in seconds to centistoke viscosity.
|
|
|
ADDITIONAL EZ CUP ADVANTAGES
EZ Cup orifices are machined rather than drilled to insure exact centering in
the cup hemisphere base and a minimum of burr formation. This insures an
orifice of specified length and a correct symmetrical efflux stream.
The EZ Cup support rods are offset from the side of the cup and secured to the
cup sidewall below the cup rim. This eliminates errors due to test material
drainage from support surfaces.
The increased separation width of the support rods by over 20% and the lowering
of weld to the cup provide best possible conditions for cleaning.
|
|
|
|
|
|